Female breast surface radiation exposure during FDG PET/CT examinations
Metin Halac1, Mehmet H Yilmaz, Kerim Sonmezoglu
1Department of Nuclear Medicine, Cerrahpasa Medical Faculty, University of Istanbul, Istanbul, Turkey. metinhallac@yahoo.com
Background:
The combined positron emission tomography (PET) and computed tomography (CT) scanners have been developed in which CT data can be used for both anatomical landmarks and attenuation correction of PET images. However, this modality potentially introduces more radiation burden to patients compared to conventional PET scanning as a result of the added radiation exposure received from CT examination. The purpose of our study was to determine the breast radiation doses of combined PET/CT examination.
Material And Methods:
Patients' superficial breast doses were calculated using thermoluminescence dosimeters (TLDs) placed onto the surface of the breasts. TLDs were positioned before FDG injection and removed after 24 h. We also determined the average superficial and glandular breast radiation doses from the anthropomorphic dosimetric phantom imaged using similar CT protocol (low dose) to the patients' study.
Results:
The mean superficial breast dose of the breast skin measured from the PET/CT studies was 14.42+/-2.41 mGy. The average superficial and glandular breast doses of the anthropomorphic phantom measured from the low-dose CT was 9.50 mGy and 5.94 mGy, respectively.
Conclusion:
This study showed that radiation exposure to the breasts during PET/CT was higher than the recommended doses. Therefore, combined PET/CT scanning must be used for essential indications, particularly in women of reproductive age and preferentially a low-dose CT protocol should be implemented to avoid overexposure in such patients.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography (FDG-PET/CT)
Published on: May 2, 2012
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT