Related Experiment Video
Updated: Jul 9, 2026

Acquisition and Semi-Automated Analysis of Respiratory Muscle Surface Electromyography
Published on: January 24, 2025
Detecting ineffective triggering in the expiratory phase in mechanically ventilated patients based on airway flow and
Chang-Wen Chen1, Wei-Chieh Lin, Chih-Hsin Hsu
1Medical Intensive Care Unit, Department of Internal Medicine, National Cheng Kung University Medical Center, Tainan, Taiwan.
Objective:
Ineffective triggering (IT) is the most common manifestation of patient-ventilator asynchrony in mechanically ventilated patients. IT in the expiratory phase (ITE) accounts for the majority of IT and is associated with characteristic features of flow and airway pressure deflection, caused by ineffective effort from the patient. The purpose of this study was to quantify the characteristics of flow and airway pressure deflections of ITE and, using a computerized algorithm, to evaluate their usefulness in the detection of ITEs.
Design:
Prospective, clinical study.
Setting:
Medical intensive care unit in a 1,000-bed university hospital.
Patients:
A total of 14 mechanically ventilated adult patients with patient-ventilator asynchrony.
Interventions:
None.
Measurements And Main Results:
We analyzed 5,899 breaths and found that 1,831 were ITEs. The average values for maximum flow deflection (F(def)) and maximum airway pressure deflection (P(def)) in ITEs were 13.94 +/- 8.0 L/min and 1.91 +/- 0.97 cm H2O. With a starting value of 0.1 L/min for F(def) and 0.01 cm H2O for P(def), the area under the receiver operating characteristics curve of F(def) and P(def) for the detection of ITEs was 0.98 and 0.97, respectively. Sensitivity and specificity for the detection of ITEs were 91.5% and 96.2% for F(def), respectively, for a cutoff value of 5.45 L/min, and 93.3% and 92.9% for Pdef, for a cutoff value of 0.45 cm H2O.
Conclusion:
We conclude that accurately detecting and quantifying ITEs is feasible using a computerized algorithm based on F(def) and P(def). Such a computerized estimation of patient-ventilator interaction might be helpful for adjusting ventilator settings in an intensive care unit.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: