Related Experiment Video
Updated: Jul 9, 2026

Comparative Analysis of Automatic Fecal Analyzer versus Direct Wet Smear Microscopy for Detecting Parasitic Infections in Stool Samples
Published on: April 25, 2025
[Evaluation of the OC-SENSOR neo system for testing fecal occult blood]
Ji Hyun Kim1, Hee-Jung Chung, Nam Surp Yoon
1Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea.
Background:
Fecal occult blood tests (FOBTs) have been widely used as a means of colorectal cancer screening. Automated FOBTs using immunologic principles have the advantages such as quantitation, high specificity, and high throughput. We evaluated a newly-introduced automated FOBT analyzer, OC-SENSOR neo (OC neo) (Eiken Chemical Co., Japan).
Methods:
The precision, linearity, and carry-over rate of OC neo were assessed with specimens prepared in accordance with the guidelines of CLSI. We performed a parallel test between OC neo and OC-SENSOR I (OC I) (Eiken Chemical Co.) using 300 consecutive stool specimens and 60 OC I-positive specimens. The results were analyzed with SPSS version 13.0 (SPSS Inc., USA).
Results:
The coefficients of variation (CV) of within-run, between-run, and between-day using OCControl L (Eiken Chemical Co.) of ca. 150 ng/mL were 3.5-7.8%, 4.5-8.8% and 4.9-5.0%, respectively. The linear regression coefficient and carry-over rate with the range of 67.8-939.4 ng/mL were 0.9998 (P<0.001), and 0.1%, respectively. Correlation coefficient between OC neo and OC I was R2=0.954 (P<0.001) for 60 OC I-positive specimens. The positive and negative interpretations of 300 consecutive specimens by OC neo were completely consistent with those of OC I.
Conclusions:
Because OC neo showed an excellent performance and a good correlation with OC I, OC neo warrants to be a reliable quantitative FOBT system for high volume laboratories.
Related Concept Videos
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Serum Laboratory Studies, Stool Test, Breath Test