Comparison between 180 degrees and 360 degrees acquisition arcs with and without correction by CT-based attenuation
Dimitris J Apostolopoulos1, Trifon Spyridonidis, Theodoros Skouras
1Department of Nuclear Medicine, University Hospital of Patras, Greece. dimap@med.upatras.gr
Objective:
To investigate the effect of acquisition arc (180 degrees or 360 degrees ) and attenuation correction (AC), in artifact occurrence and image uniformity of myocardial perfusion SPECT studies in normal hearts.
Methods:
Twenty male and 44 female patients with primary hyperparathyroidism and low probability of resting perfusion abnormalities were analysed. Tc-sestamibi myocardial perfusion studies were performed at rest with a hybrid SPECT/CT system. An anthropomorphic phantom was also imaged.
Results:
Without AC, a 360 degrees orbit resulted in significantly lower artifact rates compared to the standard 180 degrees in both gender groups. Differences persisted (P<0.001) even after adjusting reconstruction filter settings to compensate for disproportionate count statistics between the two arcs. However, gender-related variation was not considerably decreased and sub-diaphragmatic activity was augmented with a 360 degrees arc. After AC with either arc, substantially lower defect scores and minimal gender-related differences were obtained compared to the standard technique. A 360 degrees arc yielded lower defect rates and count variability than did the 180 degrees arc (P<0.05). These differences were small and were eliminated after weighting for dissimilar count statistics. Sub-diaphragmatic signal was amplified, particularly with 360 degrees ; severe bowel-myocardial activity overlap complicated image interpretation in 14% of cases with 360 degrees versus 8% with 180 degrees . In certain aspects, phantom imaging findings diverged from patient studies.
Conclusion:
Without AC there is significant disparity between 180 degrees and 360 degrees concerning artifact occurrence. After CT-based AC differences between arcs in defect rate and image uniformity seem less critical. However, 180 degrees is favoured by less intensification of intestinal activity.