Teryl K Nuckols1, Anthony G Bower, Susan M Paddock
1The RAND Corporation, Santa Monica, CA, USA. teryl@rand.org
Background:
Patients in intensive care units (ICUs) frequently experience adverse drug events involving intravenous medications (IV-ADEs), which are often preventable.
Objectives:
To determine how frequently preventable IV-ADEs in ICUs match the safety features of a programmable infusion pump with safety software ("smart pump") and to suggest potential improvements in smart-pump design.
Design:
Using retrospective medical-record review, we examined preventable IV-ADEs in ICUs before and after 2 hospitals replaced conventional pumps with smart pumps. The smart pumps alerted users when programmed to deliver duplicate infusions or continuous-infusion doses outside hospital-defined ranges.
Participants:
4,604 critically ill adults at 1 academic and 1 nonacademic hospital.
Measurements:
Preventable IV-ADEs matching smart-pump features and errors involved in preventable IV-ADEs.
Results:
Of 100 preventable IV-ADEs identified, 4 involved errors matching smart-pump features. Two occurred before and 2 after smart-pump implementation. Overall, 29% of preventable IV-ADEs involved overdoses; 37%, failures to monitor for potential problems; and 45%, failures to intervene when problems appeared. Error descriptions suggested that expanding smart pumps' capabilities might enable them to prevent more IV-ADEs.
Conclusion:
The smart pumps we evaluated are unlikely to reduce preventable IV-ADEs in ICUs because they address only 4% of them. Expanding smart-pump capabilities might prevent more IV-ADEs.
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Impact of Pharmacokinetic–Pharmacodynamic Models: Regulatory Decisions
Modified-Release Drug Delivery Systems: Rate-Programmed I