Related Experiment Video
Updated: Jul 9, 2026

Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements
Tira Bunyaviroch1, Timothy G Turkington, Terence Z Wong
1Department of Radiology, Nuclear Medicine Section, Duke University Medical Center, Durham, NC 27710, USA. tbunyaviroch@hotmail.com
Purpose:
The presence of contrast materials on computed tomography (CT) images can cause problems in the attenuation correction of positron emission tomography (PET) images. These are because of errors converting the CT attenuation of contrast to 511-keV attenuation and by the change in tissue enhancement over the duration of the PET emission scan. Newer CT-based attenuation correction (CTAC) algorithms have been developed to reduce these errors.
Methods:
To evaluate the effectiveness of the modified CTAC technique, we performed a retrospective analysis on 20 patients, comparing PET images using unenhanced and contrast-enhanced CT scans for attenuation correction. A phantom study was performed to simulate the effects of contrast on radiotracer concentration measurements.
Results:
There was a maximum difference in calculated radiotracer concentrations of 5.9% within the retrospective data and 7% within the phantom data.
Conclusion:
Using a CTAC algorithm that de-emphasizes high-density areas, contrast-enhanced CT can be used for attenuation mapping without significant errors in quantitation.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
![Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F58491.jpg&w=3840&q=50)