Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography
Ulrike Ropers1, Dieter Ropers, Tobias Pflederer
1Department of Internal Medicine 2 (Cardiology-Angiology), University of Erlangen-Nuremberg, Erlangen, Germany.
Insights
Dual-source CT coronary angiography shows slightly reduced image quality at higher heart rates but maintains diagnostic accuracy for detecting coronary artery stenosis. This study evaluated 100 patients without beta-blocker pre-medication.
Area of Science:
- Cardiovascular Imaging
- Radiology
- Medical Diagnostics
Background:
- Multidetector CT (MDCT) coronary angiography has shown an inverse relationship between heart rate and image quality.
- Dual-source CT (DSCT) offers higher temporal resolution, potentially improving image quality at elevated heart rates.
Purpose of the Study:
- To assess the impact of heart rate on image quality and diagnostic accuracy of DSCT coronary angiography.
- To determine if higher heart rates affect the ability to detect coronary artery stenoses using DSCT.
Main Methods:
- One hundred patients underwent DSCT coronary angiography with a contrast-enhanced volume dataset acquisition.
- Image datasets were evaluated for significant coronary stenoses and validated against invasive coronary angiography.
- Patients were analyzed based on heart rate: >=65 beats/min versus <65 beats/min, without beta-blocker pre-medication.
Main Results:
- Per-segment evaluability was 92% for heart rates >=65 bpm and 100% for heart rates <65 bpm (p < 0.001).
- Per-patient evaluability was 93% for heart rates >=65 bpm and 100% for heart rates <65 bpm.
- Despite slightly lower evaluability at higher heart rates, diagnostic accuracy (sensitivity, specificity) for detecting coronary stenoses did not differ significantly between heart rate groups.
Conclusions:
- DSCT coronary angiography demonstrates slightly reduced per-segment evaluability in patients with heart rates >=65 bpm.
- However, DSCT maintains diagnostic accuracy for detecting coronary artery stenoses even at higher heart rates without beta-blocker pre-medication.
Objectives:
We evaluated the influence of heart rate on image quality and diagnostic accuracy of dual-source computed tomography (DSCT) coronary angiography.
Background:
Multidetector computed tomography (MDCT) coronary angiography has demonstrated an inverse relationship between heart rate and image quality. Dual-source CT provides a higher temporal resolution.
Methods:
One hundred patients were studied by DSCT (DEFINITION, Siemens Medical Solutions, Forchheim, Germany). A contrast-enhanced volume dataset was acquired (two tubes, 120 kV, 400 mAs/rot, collimation 64 x 0.6 mm). Datasets were evaluated concerning the presence of significant coronary stenoses and validated against invasive coronary angiography.
Results:
In 44 patients with a heart rate > or =65 beats/min, 566 of 616 coronary segments were evaluable (92%), whereas in 56 patients with a heart rate <65 beats/min, 777 of 778 coronary segments were evaluable (100%, p < 0.001). On a per-patient basis, 93% of patients (> or =65 beats/min) and 100% of patients (<65 beats/min) were considered evaluable. By classifying unevaluable segments as positive for stenosis, per-patient sensitivity was 95% (19 of 20) for heart rates > or =65 beats/min and 100% (22 of 22) for heart rates <65 beats/min. Specificity was 87% (21 of 24) versus 76% (26 of 34), and overall diagnostic accuracy was 91% (40 of 44) versus 86% (48 of 56). None of these differences were statistically significant. Similarly, no difference in diagnostic accuracy was found in per-vessel and -segment analyses.
Conclusions:
In 100 patients studied without beta-blocker pre-medication, DSCT demonstrated slightly lower per-segment evaluability for high heart rates but no decrease in diagnostic accuracy for the detection of coronary artery stenoses.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...