Enhanced photoemission spectroscopy for verification of high-risk i.v. medications
Yaman Kaakeh1, Hanna Phan, Brian D DeSmet
1Purdue University, West Lafayette, IN, USA.
Purpose:
The sensitivity and specificity of enhanced photoemission spectroscopy (EPS) for performing an automated final check of compounded i.v. admixtures at a pediatric hospital pharmacy were studied.
Methods:
A tabletop EPS device was used to test samples of seven high-risk drug-diluent combinations compounded in the pharmacy; the drugs were vancomycin, lorazepam, morphine, insulin, hydromorphone, gentamicin, and epinephrine. Ten sets of samples were prepared for each drug. Typically, a sample set consisted of dilutions ranging from 10-fold above to 10-fold below the targeted concentration. Testing was performed twice weekly between November 2005 and March 2006.
Results:
The EPS device detected errors departing from the targeted concentration by 20% or more with a sensitivity of at least 95%. Specificity in distinguishing among test medications at targeted concentrations was 100%. The percentage of passing samples with intermediate concentrations varied among the drugs.
Conclusion:
A tabletop EPS device demonstrated acceptable sensitivity and specificity for validating the identity and concentrations of selected high-risk i.v. medications compounded for pediatric patients. The device may help prevent clinically important medication errors caused by inaccurate compounding.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Atomic Emission Spectroscopy: Lab
UV–Vis Spectrometers
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Interference