Related Experiment Video
Updated: Jul 8, 2026

Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
Coronary artery spasm as a cause of ST elevation and inappropriate implantable cardioverter defibrillator
Giosuè Mascioli1, Luca Bontempi, Marco Racheli
1Department of Cardiology, Spedali Civili di Brescia, Italy. giosue.mascioli@virgilio.it
Insights
Coronary artery spasm can cause arrhythmias and trigger inappropriate implantable cardioverter defibrillator (ICD) firing due to ST segment elevation. This case highlights a unique ICD malfunction caused by coronary artery spasm.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Coronary artery spasm is a known cause of cardiac arrhythmias, including bradycardia and tachyarrhythmia.
- Electrocardiographically, coronary artery spasm typically manifests as ST segment elevation.
- Implantable cardioverter defibrillators (ICDs) are used to manage life-threatening ventricular arrhythmias.
Observation:
- A patient with an ICD for unexplained syncope experienced coronary artery spasm.
- During the spasm, the ICD delivered an inappropriate shock.
- The inappropriate shock was attributed to ST segment elevation mimicking a ventricular arrhythmia.
Findings:
- The ST segment elevation caused by coronary artery spasm led to double counting of the QRS complex by the ICD.
- This QRS counting error resulted in the device misinterpreting the rhythm as a life-threatening arrhythmia.
- Consequently, the ICD delivered an inappropriate shock.
Implications:
- This case underscores the potential for coronary artery spasm to cause ICD malfunction.
- Accurate interpretation of ECG changes during coronary artery spasm is crucial for appropriate ICD programming and function.
- Further research may be needed to refine ICD algorithms to differentiate between true arrhythmias and artifact from coronary artery spasm.
Abstract:
Coronary artery spasm can cause both brady- and tachyarrhythmia, through induction of AV block (usually linked to coronary spasm of the right coronary artery) or ventricular tachycardia/fibrillation linked to extensive myocardial ischemia. The electrocardiographic aspect of coronary artery spasm is an ST segment elevation. We describe the case of patient implanted with an implantable cardioverter defibrillator (ICD) for unexplained syncope which, during coronary artery spasm, received an inappropriate device firing due to ST segment elevation, leading to a double count of the QRS by the ICD.
More Related Videos
06:39Intracoronary Acetylcholine Provocation Testing for Assessment of Coronary Vasomotor Disorders
Published on: August 18, 2016
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Acute Coronary Syndrome I: Introduction
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Acute Coronary Syndrome IV: Interprofessional Care
Coronary Artery Disease V: Interprofessional Care
Acute Coronary Syndrome III: Diagnostic Studies
Cardiomyopathy II: Dilated Cardiomyopathy