Related Experiment Video
Updated: Jul 8, 2026

Induction of Microstreaming by Nonspherical Bubble Oscillations in an Acoustic Levitation System
Published on: May 9, 2021
Design and implementation of an efficient acoustically levitated drop reactor for in stillo measurements
Christopher R Field1, Alexander Scheeline
1Department of Chemistry, University of Illinois Urbana-Champaign, 600 S. Mathews Ave. Urbana, Illinois 61801, USA. cfield2@scs.uiuc.edu
Abstract:
We present the details necessary for building an efficient acoustic drop levitator with reduced electrical power consumption and greater drop stability compared to previous designs. The system is optimized so that the levitated drop may be used as a chemical reactor. By introducing a temperature, pressure, and relative humidity sensor for feedback control of a linear actuator for adjusting resonator length, we have built a completely automated system capable of continuous levitation for extended periods of time. The result is a system capable of portable operation and interfacing with a variety of detection instrumentation for in stillo (in drop) measurements.