Combined T1-based perfusion MRI and MR angiography in kidney: first experience in normals and pathology
Martine Dujardin1, Rob Luypaert, F Vandenbroucke
1Department of Radiology/BEFY, UZ Brussel, Vrije Universiteit Brussel, Laarbeeklaan 101, 1090 Brussels, Belgium. martine.dujardin@gmail.com
Objectives:
To investigate the feasibility of implementing quantitative T1-perfusion in the routine MRA-protocol and to obtain a first experience in normals and pathology.
Materials And Methods:
For perfusion imaging, IR-prepared FLASH (one 4 mm slice at mid-renal level, TR 4.4 ms, TE 2.2 ms, TI 180 ms, FA 50 degrees , matrix 128 x 256, bandwidth per pixel 300, 400 dynamics, temporal resolution 0.3 s, total measurement time 2 min) was applied during the injection of 10 ml of standard 0.5 mmol/ml Gadolinium-DTPA solution at 2 ml/s, followed by 3DCE-MRA with bolus tracking (TR 5.4, TE 1.4, FA 40 degrees , matrix 192 x 512, NSA 1, slice thickness 1.5 mm), using a second dose of 0.1 mmol Gadolinium-DTPA per kg body weight with a maximum of 20 ml. The T1-weighted signals (perfusion data) were converted to tissue tracer concentrations and deconvolved with an inflow corrected AIF; blood flow, distribution volume, mean transit time and blood flow heterogeneity were derived.
Results:
MRA quality was uncompromised by the first bolus administered for perfusion purposes. In the normals, average cortical RBF, RVD and MTT were 1.2 ml/min/ml (S.D. 0.3 ml/min/ml), 0.4 ml/ml (S.D. 0.1 ml/ml) and 21s (S.D. 4s). These RBF values are lower than those found in the literature, probably due to residual AIF inflow effects. The sensitivity of the technique was sufficient to demonstrate altered perfusion in the examples of pathology.
Conclusion:
Combined quantitative T1-perfusion and MRA have a potential for noninvasive renovascular screening and may provide an anatomical and physiological evaluation of renal status.
More Related Videos
10:16Retrograde Perfusion and Filling of Mouse Coronary Vasculature as Preparation for Micro Computed Tomography Imaging
Published on: February 10, 2012
06:02Anatomical Reconstructions of the Human Cardiac Venous System using Contrast-computed Tomography of Perfusion-fixed Specimens
Published on: April 18, 2013
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies for Cardiovascular System V: CT