Related Experiment Video
Updated: Jul 8, 2026

10:25
Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025
Projector-based augmented reality for intuitive intraoperative guidance in image-guided 3D interstitial brachytherapy
Robert Krempien1, Harald Hoppe, Lüder Kahrs
1Department of Radiation Oncology, University of Heidelberg, Heidelberg, Germany. robert_krempien@med.uni-heidelberg-de
International Journal of Radiation Oncology, Biology, Physics
|January 1, 2008
Summary
This study introduces an augmented reality system for real-time image-guided interstitial brachytherapy, enhancing surgeon orientation during procedures. The low-cost system proved accurate and feasible for needle implantation monitoring.
Area of Science:
- Medical Imaging and Image-Guided Therapy
- Augmented Reality in Medicine
- Radiation Oncology
Background:
- Interstitial brachytherapy requires precise needle placement for effective radiation delivery.
- Intraoperative orientation can be challenging, impacting surgical accuracy and efficiency.
- Real-time image guidance systems aim to improve surgical outcomes in brachytherapy.
Purpose of the Study:
- To implement an augmented reality (AR) system for real-time image guidance in interstitial brachytherapy.
- To provide intuitive intraoperative orientation for surgeons during brachytherapy procedures.
- To evaluate the accuracy and feasibility of the developed AR system in a clinical setting.
Main Methods:
- A low-cost AR system comprising a video projector, two cameras, and a notebook was developed.
- Coded-light patterns projected by the system registered the patient and superimposed planning data onto the surgical field.
- Stereoscopic tracking of passive markers on the patient detected movements, allowing dynamic adjustment of projected data without rigid fixation.
Main Results:
- The system enabled surgeons to visualize planning data directly on the patient's surface with a clear line of sight.
- The tracking system compensated for patient movement (average deviation 1.1 mm) and projector repositioning (average deviation 0.9 mm).
- Mean deviation for implanted needles was 1.4 mm, demonstrating high accuracy in a clinical study of 10 patients.
Conclusions:
- The developed low-cost augmented reality system is accurate and feasible for interstitial brachytherapy.
- The system provides intuitive real-time intraoperative orientation and monitoring of needle implantation.
- The AR system meets clinical demands for improved image-guided brachytherapy.