Related Experiment Video
Updated: Jul 8, 2026

Manufacture of a Multi-Purpose Low-Cost Animal Bench-Model for Teaching Tracheostomy
Published on: May 18, 2019
A comparison of two emergency cricothyroidotomy kits in human cadavers
Mehdi Benkhadra1, François Lenfant, Wolfgang Nemetz
1Division of Anatomy, University of Geneva, Geneva, Switzerland. m.benkhadra@wanadoo.fr
Background:
We compared two emergency cricothyroidotomy kits designed to avoid lesions during insertion, one based on the Seldinger technique (ST), the other based on the concept of a mechanical detection of the posterior wall of the larynx, with regard to insertion time, success rate, and complication rate.
Methods:
Cricothyroidotomy was performed under fiberoptic control in 40 human cadavers embalmed according to Thiel's technique. The set chosen for use was randomized: new technique (NT) or ST. Duration of the procedure, success rates, and incidence of laryngeal injuries were compared. Traumatic lesions observed with the fiberoptic bronchoscope were anatomically confirmed after dissection.
Results:
The two groups had comparable epidemiological and anatomical records. Cricothyroidotomy was performed faster with the NT than with the ST (median 54 vs 71 s, P = 0.01). Failure rates were comparable between groups (4 vs 1, P = 0.34), and there were fewer major complications in the posterior tracheal wall with the ST (0 vs 8, P = 0.003). In the ST group, only minor punctiform lesions of the posterior trachea wall were observed in four cases.
Conclusions:
In this model, despite a shorter insertion time, the NT produced more lesions and more failures than the ST.
More Related Videos
05:25Identification and Protection of the Recurrent Laryngeal Nerve during Transoral Robotic Thyroidectomy
Published on: October 24, 2025
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022