PHOTOMETER FOR MEASURING THE SCOTOPIC CANDLEPOWERS OF SELF-LUMINOUS OPHTHALMIC TEST OBJECTS
1The National Physical Laboratory.
The British Journal of Ophthalmology
|December 1, 1944
Summary
This study introduces a new photometer capable of measuring extremely low light levels down to 10(-9) candle. This instrument is crucial for calibrating radium paint light sources used in mapping the dark-adapted visual field.
Area of Science:
- Photometry
- Ophthalmology
- Radiological Physics
Background:
- Measuring feeble light sources is essential for visual field testing.
- Radium paint spots are used for plotting the dark-adapted visual field.
- Existing photometers may lack the sensitivity for these applications.
Purpose of the Study:
- To describe a novel photometer for measuring scotopic candlepowers of very feeble light sources.
- To calibrate small radium paint spots used in visual field studies.
- To validate the photometer's adherence to the scotopic visibility curve.
Main Methods:
- Development of a sensitive photometer.
- Calibration of radium paint sources (2-3 mm diameter) with candlepowers in the range of 10(-9) candle.
- Testing with color filters and spectral transmission curves to verify scotopic measurement.
Main Results:
- The photometer successfully measures scotopic candlepowers down to approximately 10(-9) candle.
- The instrument is suitable for calibrating radium paint light sources.
- Measurements align with the scotopic visibility curve of the human eye.
Conclusions:
- The described photometer provides accurate measurements of feeble light sources.
- It is a valuable tool for calibrating instruments used in ophthalmological research.
- The photometer's performance is validated against the standard scotopic visibility curve.
Related Concept Videos
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
Spectrophotometry: Introduction
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Flame Photometry: Overview
Flame photometry, also known as flame emission spectrometry, is a technique used for the qualitative and quantitative analysis of elements present in a sample using a flame as the source of excitation energy. The concept of flame photometry was realized in the early 1860s by Kirchhoff and Bunsen, who discovered that specific elements emit characteristic radiation when excited in flames. The first instrument developed for this purpose was used to measure sodium (Na) in plant ash using a Bunsen...