Sensing failure associated with the Medtronic Sprint Fidelis defibrillator lead
David N Kenigsberg1, Sunil Mirchandani, Amanda N Dover
1Virginia Commonwealth University Medical Center, Richmond, Virginia 23298-0053, USA.
Introduction:
The diameter of implantable cardioverter-defibrillator (ICD) leads has become progressively smaller over time. However, the long-term performance characteristics of these smaller ICD leads are unknown.
Methods:
We retrospectively evaluated 357 patients who underwent implantation of a Medtronic Sprint Fidelis defibrillating lead at two separate centers between September 2004 and October 2006. Lead characteristics were measured at implant, at early follow-up (1-4 days post implant), and every 3-6 months thereafter.
Results:
During the study period, 357 patients underwent implantation of the Medtronic Sprint Fidelis lead. The mean R-wave measured at implant through the device was not different (P = NS) when compared with that measured at first follow-up (10.5 +/- 5.0 mV vs 10.7 +/- 5.1 mV). Forty-one patients (13%) had an R-wave amplitude
Conclusion:
Abnormal R-wave sensing is frequently observed during follow-up with the Medtronic Fidelis ICD lead. Lead revision was necessary in 2.8% of the patients, most often (8 of 10) due to abnormal R-wave sensing along with elevated pacing threshold. Whether this issue is limited to this lead or reflects a potential problem with all downsized ICD leads merits further investigation.
More Related Videos
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Errors occurring during blood pressure monitoring
Several factors...
Cardiomyopathy II: Dilated Cardiomyopathy
Dysrhythmias V: Evaluating Dysrhythmias