Influence of the polymerization-mixture composition for monolithic methacrylate-based columns on the
Indiana Tanret1, Debby Mangelings, Yvan Vander Heyden
1Department of Analytical Chemistry and Pharmaceutical Technology, Pharmaceutical Institute, Vrije Universiteit Brussel-VUB, Laarbeeklaan 103, B-1090 Brussels, Belgium.
Abstract:
Methacrylate-based monolithic stationary phases were evaluated for the analysis of drug molecules using capillary electrochromatography (CEC) as separation technique. The effect of the polymerization-mixture composition on the retention behavior of a small test set of mainly drug molecules was studied. Two factors were varied in a central-composite design-based approach: the ratio between the pore-forming solvents and the monomers on one hand, and the ratio within the pore-forming solvents on the other hand, resulting in nine different stationary phases. The central point of the design was chosen at 70% (m/m) pore-forming solvents (PFS) of which 30% (m/m) is 1,4-butanediol, i.e. 21% of the total polymerization mixture. Experiments were conducted using both a basic (pH 11.5) and an acidic (pH 3) mobile phase. Retention times, retention factors, peak asymmetry and number of theoretical plates are the responses used to evaluate the performance of the resulting monoliths. The best compromise between the different responses was found around 67% PFS and 18% 1,4-butanediol (relative to the total mass), i.e. rather close to the center point. At these conditions, retention times were generally below 15min and retention factors below 5. Asymmetry values between close to 1 were found, and theoretical plate numbers up to 10,900, which were improvements compared to the central point of the design.
Related Concept Videos
Principles Of Column Chromatography
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Determination of Molar Masses of Polymers II
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...