Related Experiment Video
Updated: Jul 8, 2026

Dual-mode Imaging of Cutaneous Tissue Oxygenation and Vascular Function
Published on: December 8, 2010
A sensitivity analysis of the step-temperature technique for measurement of local tissue blood perfusion
1College of Energy and Power Engineering, Huazhong University of Science and Technology, Wuhan, Hubei, PR China. yk_hust@163.com
Abstract:
The step-temperature technique is a thermal technique that uses a small thermistor probe for measuring blood perfusion of tissue. The blood perfusion is derived from temperature and power measurements using equations that describe heat transfer in the integrated probe/tissue system. A numerical experiment is used to analyse the theoretical error caused by the assumptions used in the technique. The effects of the bead parameters, tissue parameters and measurement parameters are investigated. The study results are in accordance with experimental phenomena described previously. An optimal measurement time window is found to be 4-10 s, dependent mainly on the thermal conductivity and blood perfusion of tissue measured. This research will lead to reduce perfusion measurement errors.
More Related Videos
10:23A Computational Modeling Approach to Investigate the Influence of Hyperthermia on the Tumor Microenvironment
Published on: December 1, 2023
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024