Assessment of stroke volume index with three different bioimpedance algorithms: lack of agreement compared to
Eric E C de Waal1, Maurits K Konings, Cor J Kalkman
1Division of Perioperative and Emergency Care, University Medical Center, Heidelberglaan 100, 3584 CX, Utrecht, The Netherlands. e.e.c.dewaal@azu.nl
Objective:
The accuracy of bioimpedance stroke volume index (SVI) is questionable as studies report inconsistent results. It remains unclear whether the algorithms alone are responsible for these findings. We analyzed the raw impedance data with three algorithms and compared bioimpedance SVI to transpulmonary thermodilution (SVI(TD)).
Design And Setting:
Prospective observational clinical study in a university hospital.
Patients:
Twenty adult patients scheduled for coronary artery bypass grafting (CABG).
Interventions:
SVI(TD) and bioimpedance parameters were simultaneously obtained before surgery (t1), after bypass (t2), after sternal closure (t3), at the intensive care unit (t4), at normothermia (t5), after extubation (t6) and before discharge (t7). Bioimpedance data were analyzed off-line using cylinder (Kubicek: SVI(K); Wang: SVI(W)) and truncated cone based algorithms (Sramek-Bernstein: SVI(SB)).
Measurements And Results:
Bias and precision between the SVI(TD) and SVI(K), SVI(SB), and SVI(W) was 1.0+/-10.8, 9.8+/-11.4, and -15.7+/-8.2 ml/m2 respectively, while the mean error was abundantly above 30%. Analysis of data per time moment resulted in a mean error above 30%, except for SVI(W) at t2 (28%).
Conclusions:
Estimation of SVI by cylinder or truncated cone based algorithms is not reliable for clinical decision making in patients undergoing CABG surgery. A more robust approach for estimating bioimpedance based SVI may exclude inconsistencies in the underlying algorithms in existing thoracic bioimpedance cardiography devices.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Assessment of blood pressure in brachial artery(two-step method)