Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Comparison of optic disc margin identified by color disc photography and high-speed ultrahigh-resolution optical
Anita Manassakorn1, Hiroshi Ishikawa, Jong S Kim
1UPMC Eye Center, Eye and Ear Institute, Department of Ophthalmology, University of Pittsburgh School of Medicine, Pittsburgh, PA 15213, USA.
Objective:
To determine the correspondence between optic disc margins evaluated using disc photography (DP) and optical coherence tomography (OCT).
Methods:
From May 1, 2005, through November 10, 2005, 17 healthy volunteers (17 eyes) had raster scans (180 frames, 501 samplings per frame) centered on the optic disc taken with stereo-optic DP and high-speed ultrahigh-resolution OCT (hsUHR-OCT). Two image outputs were derived from the hsUHR-OCT data set: an en face hsUHR-OCT fundus image and a set of 180 frames of cross-sectional images. Three ophthalmologists independently and in a masked, randomized fashion marked the disc margin on the DP, hsUHR-OCT fundus, and cross-sectional images using custom software. Disc size (area and horizontal and vertical diameters) and location of the geometric disc center were compared among the 3 types of images.
Results:
The hsUHR-OCT fundus image definition showed a significantly smaller disc size than the DP definition (P <.001, mixed-effects analysis). The hsUHR-OCT cross-sectional image definition showed a significantly larger disc size than the DP definition (P <.001). The geometric disc center location was similar among the 3 types of images except for the y-coordinate, which was significantly smaller in the hsUHR-OCT fundus images than in the DP images.
Conclusion:
The optic disc margin as defined by hsUHR-OCT was significantly different than the margin defined by DP.