Fast and Accurate Exhaled Breath Ammonia Measurement
Published on: June 11, 2014
A laboratory intercomparison of real-time gaseous ammonia measurement methods
James J Schwab1, Yongquan Li, Min-Suk Bae
1Atmospheric Sciences Research Center, University at Albany, State University of New York, USA. schwab@asrc.cestm.albany.edu
Abstract:
Six different measurement methods (and seven instruments) for the measurement of gaseous ammonia at low part-per-billion levels were compared simultaneously in a laboratory setting. The instruments were the tunable diode laser (TDL) absorption spectrometer, the wet scrubbing long-path absorption photometer (LOPAP), the wet effusive diffusion denuder (WEDD), the ion mobility spectrometer (IMS), the Nitrolux laser acousto-optical absorption analyzer, and a modified chemiluminescence analyzer. With the exception of the modified chemiluminescence analyzer, the instruments performed well and, understable calibration conditions, generally agreed to within about 25% of the expected calibration value. Instrument time response is shown to be sensitive to measurement history as well as the sample handling materials and is shortest for the TDL. The IMS and Nitrolux are commercial instruments used without modification from the manufacturer. These two instruments have significantly slower time response than the TDL (especially in the case of the Nitrolux) and exhibited measurement biases of approximately +25% (IMS) and -25% (Nitrolux). The LOPAP and WEDD instruments, both research instruments using wet chemical methods, performed well in the calibration tests in terms of the absolute accuracy of measured concentrations, but the WEDD instrument suffered from significantly slower time response than the LOPAP.
More Related Videos
10:29Calibrated Passive Sampling - Multi-plot Field Measurements of NH3 Emissions with a Combination of Dynamic Tube Method and Passive Samplers
Published on: March 21, 2016
08:05Measurement of the Potential Rates of Dissimilatory Nitrate Reduction to Ammonium Based on 14NH4+/15NH4+ Analyses via Sequential Conversion to N2O
Published on: October 7, 2020
Related Concept Videos
Chemical Stoichiometry and Gases: Using Ideal Gas Law to Determine Moles
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Gas Chromatography: Types of Detectors-II