Development of a reflected optical fiber system for measuring oxygen saturation in an integrated artificial
Toshitaka Yasuda1, Tomohiko Saito, Tatsuya Kihara
1Graduate School of Medicine, Yamaguchi University, Ube, Yamaguchi, Japan. tyasuda@yamaguchi-u.ac.jp
Abstract:
The purpose of this study was to develop a blood oxygen saturation (OS) monitoring system for use with an integrated artificial heart-lung system (IAHLS). The OS monitoring system consists of two paired optical fiber probes (OFPs) and a measurement system. To investigate the effect of the OFP configuration and incident light wavelength on the relationship between OS and the reflectance ratio for wavelengths of 810 and 645 nm, we performed theoretical analyses of the relationship between OS and R810/R645 using a diffusion equation. The prototype OFP located on the blood outlet port of our IAHLS housing was evaluated using an in vitro test. An OS range of 65-100% was adjusted to supply oxygen and nitrogen gas to the IAHLS. The blood flow rate was maintained at 3 L/min by the rotational speed of an impeller in the IAHLS. The OS-corrected blood from the IAHLS was measured using a commercial gas analyzer. The correlation coefficients (r(2)) between the theoretical ratio of R810/R645 and OS, and between measured OS and the reflectance ratio of R810/R645 were 0.97 and 0.78, respectively. In conclusion, we confirmed that the development of this oximetry system is applicable for IAHLS.
More Related Videos
08:51Intra-cardiac Side-Firing Light Catheter for Monitoring Cellular Metabolism using Transmural Absorbance Spectroscopy of Perfused Mammalian Hearts
Published on: May 12, 2019
06:26Maintaining and Assessing Various Tissue and Cell Types of the Eye Using a Novel Pumpless Fluidics System
Published on: July 14, 2023
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...