Magnification characteristics of the Optical Coherence Tomograph STRATUS OCT 3000
Ana Sanchez-Cano1, Begoña Baraibar, Luis E Pablo
1Department of Ophthalmology, Miguel Servet University Hospital, Instituto Aragonés de Ciencias de la Salud, Zaragoza, Spain. anasanchez3@auna.com
Purpose:
To evaluate the magnification characteristics of an Optical Coherence Tomograph (STRATUS OCT 3000) used to measure optic disc size.
Methods:
An eye model was designed, built and used to measure changes in the magnification of the Optical Coherence Tomograph (OCT) with variation in the axial length of the eye and OCT correction lenses. Theoretical calculations using ray-tracing were used to compare the experimental and theoretical results.
Results:
OCT-performed fundus measurements have a magnification that depends on two factors: eye axial length and optical correction with the focusing dial of the OCT. Theoretical calculations showed that the size of the optic disc image for greater axial lengths of the eye model (myopic eye) was less than that for smaller axial lengths (hyperopic eye). The disc diameter measurements of the image obtained with the OCT did not agree with the real disc size. By focussing with the focusing dial of the OCT, the power of the optics can be varied from -12.0 D to +12.0 D and the image size is decreased for all axial lengths. The maximum variation of the disc area was 2.1%. Differences in area measured at intervals of 2.0 D after correcting the eye refraction for measurements with the OCT were equivalent to the repeatability error at that point.
Conclusions:
Our results confirm that any variation in the optical system of the eye, in the OCT, and/or in the distance between them, induces a change in the magnification of the tomograph images. These variations will affect the measurements of any of the structures in the retinal plane.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography