An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
[Biomechanical comparison of self-drilling and self-tapping orthodontic micro-implants]
Yan Chen1, Wen-ting Zhao, Hee-moon Kyung
1Department of Orthodontics, School of Dentistry, Kyungpook National University, Daegu 700-412, Korea.
Objective:
To evaluate the biomechanical influences of self-drilling and self-tapping methods on orthodontic micro-implants under immediate loading.
Methods:
Fifty-six micro-implants placed at the interradicular areas of buccal side of upper and lower jaws in two dogs were divided into self-drilling and self-tapping groups. A horizontal load of 1.96 N was immediately given for 9 weeks. Peak insertion torque and peak removal torque were tested immediately after implantation and after the dogs were killed, respectively.
Results:
Both in the upper and lower jaws peak insertion torque was significantly higher in self-drilling group [(5.6 +/- 1.1) N.cm and (8.7 +/- 2.3) N.cm] than in self-tapping group [(3.5 +/- 2.1) N.cm and (7.4 +/- 1.1) N.cm] (P < 0.05). Both in the upper and lower jaws peak removal torque was higher in self-drilling group [(-6.5 +/- 2.2) N.cm and (-7.1 +/- 2.0) N.cm] than in self-tapping group [(-5.7 +/- 2.3) N.cm and (-6.1 +/- 0.5) N.cm]. No significant difference was found (P > 0.05). The tendency of fracture was found in self-drilling group.
Conclusions:
Self-drilling micro-implants have high initial stability and can be used in the maxilla and at thin cortical bone area of mandible.