Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Modelling of formic acid production using solventless trialkylamine reduction strategies.

Physical chemistry chemical physics : PCCP·2025
Same author

Effect of personalized spinal profile on its biomechanical response in an EMG-assisted optimization musculoskeletal model of the trunk.

Journal of biomechanics·2023
Same author

Comparison of sedation with dexmedetomidine/atipamezole administered subcutaneously at GV20 acupuncture point with usual routes of administration in dogs presented for orthopaedic radiographs.

The Journal of small animal practice·2023
Same author

Subjective self-assessment of physical activity is negatively affected by monitoring awareness in subjects with mild cognitive impairment: a crossover randomised controlled trial.

European review for medical and pharmacological sciences·2022
Same author

The density of calretinin striatal interneurons is decreased in 6-OHDA-lesioned mice.

Brain structure & function·2021
Same author

Usability testing of MySkinSelfie: a mobile phone application for skin self-monitoring.

Clinical and experimental dermatology·2019

Related Experiment Video

Updated: Jul 8, 2026

Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
06:55

Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses

Published on: June 6, 2017

Maximum likelihood positioning in the scintillation camera using depth of interaction.

D Gagnon1, N Pouliot, L Laperriere

  • 1Biomed. Eng., Montreal Heart Inst., Que.

IEEE Transactions on Medical Imaging
|January 1, 1993
PubMed
Summary

The depth of interaction (DOI) in Auger gamma cameras significantly impacts performance, causing positional and energy discrepancies. Accounting for DOI effects is crucial for accurate image formation in medical imaging.

More Related Videos

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
11:57

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material

Published on: May 20, 2013

Three-dimensional Super Resolution Microscopy of F-actin Filaments by Interferometric PhotoActivated Localization Microscopy (iPALM)
11:57

Three-dimensional Super Resolution Microscopy of F-actin Filaments by Interferometric PhotoActivated Localization Microscopy (iPALM)

Published on: December 1, 2016

Related Experiment Videos

Last Updated: Jul 8, 2026

Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
06:55

Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses

Published on: June 6, 2017

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
11:57

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material

Published on: May 20, 2013

Three-dimensional Super Resolution Microscopy of F-actin Filaments by Interferometric PhotoActivated Localization Microscopy (iPALM)
11:57

Three-dimensional Super Resolution Microscopy of F-actin Filaments by Interferometric PhotoActivated Localization Microscopy (iPALM)

Published on: December 1, 2016

Area of Science:

  • Medical Physics
  • Nuclear Medicine Imaging
  • Instrumentation Science

Background:

  • Photomultiplier (PM) response in gamma cameras is influenced by the depth of interaction (DOI).
  • Accurate quantification of DOI effects is essential for improving gamma camera performance.

Purpose of the Study:

  • To quantify the specific effects of DOI on the PM response in an Auger gamma camera.
  • To evaluate the impact of DOI on positional and energy resolution.
  • To assess the influence of non-perpendicular photon incidence on DOI-related distortions.

Main Methods:

  • Implementation and testing of a Monte Carlo simulator with detailed noise modeling.
  • Development of two comparative models: one geometric-only and one realistic camera environment.
  • Simulation of 140-keV photon interactions within a typical camera configuration and with varying obliquity angles.

Main Results:

  • DOI alone can cause a 6.4-mm positional and 12% energy discrepancy for 140-keV photons.
  • Non-perpendicular photon incidence, such as with slant hole or cone beam collimators, exacerbates DOI-induced distortions.
  • A 30-degree slant angle with a 0.95-cm crystal resulted in a 5.5-mm positional variation due to obliquity.

Conclusions:

  • Both geometrical and stochastic effects of DOI significantly degrade gamma camera performance.
  • DOI-related effects must be incorporated into the image formation process for improved accuracy.
  • Accurate modeling of DOI is critical for advancing gamma camera technology and diagnostic capabilities.