Related Concept Videos

Positron Emission Tomography01:29

Positron Emission Tomography

Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Evaluation of an MR-compatible blood sampler for PET.

Physics in medicine and biology·2010
Same author

Multi-modality imaging of uveal melanomas using combined PET/CT, high-resolution PET and MR imaging.

Nuklearmedizin. Nuclear medicine·2008
Same author

Imaging of acetylcholine esterase activity in brainstem nuclei involved in regulation of sleep and wakefulness.

European journal of neurology·2007
Same author

Decompressive hemicraniectomy in a new cat model. Methodological description of the PET study protocol.

Brain research. Brain research protocols·2004
Same author

Imaging in gene therapy of patients with glioma.

Journal of neuro-oncology·2003
Same author

A new animal model of cerebral venous infarction: ligation of the posterior part of the superior sagittal sinus in the cat.

Swiss medical weekly·2003

Related Experiment Video

Updated: Jul 8, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
11:09

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals

Published on: December 16, 2022

Scatter correction in 3-D PET.

M J Lercher1, K Wienhard

  • 1Max-Planck-Inst. fur Neurologische Forschung, Koln.

IEEE Transactions on Medical Imaging
|January 1, 1994
PubMed
Summary
This summary is machine-generated.

Scattered radiation in 3D positron emission tomography (PET) brain scans was accurately modeled. Two fast scatter correction algorithms were developed, reducing scatter fractions to 5% or less for improved image quality.

More Related Videos

Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography
09:41

Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography

Published on: January 20, 2023

Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

Related Experiment Videos

Last Updated: Jul 8, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
11:09

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals

Published on: December 16, 2022

Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography
09:41

Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography

Published on: January 20, 2023

Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Physics

Background:

  • Modern multiring positron emission tomographs (PET) acquire 3D data for increased sensitivity.
  • Scattered radiation constitutes a significant portion of acquired PET data, potentially degrading image quality.
  • Accurate scatter correction is crucial for quantitative analysis in 3D PET imaging, especially for brain studies.

Purpose of the Study:

  • To experimentally characterize point source scatter distributions in 3D PET relevant to brain imaging.
  • To develop and validate fast scatter correction algorithms for 3D PET brain scans.
  • To assess the effectiveness of scatter correction in reducing scatter fractions.

Main Methods:

  • Experimental measurement of point source scatter distributions varying energy window, source location, and scatter volume.
  • Parametrization of scatter distributions using a 2D Gaussian with a shift parameter.
  • Formulation of two fast 2D-based scatter correction algorithms for 3D PET data.
  • Implementation and testing of algorithms with attenuation correction on point source and phantom data.

Main Results:

  • Point source scatter distributions were accurately parametrized by a 2D Gaussian model.
  • Developed algorithms effectively transformed 2D projection subsets into scatter projections for reconstruction.
  • Reconstructed scatter fractions were significantly reduced, achieving 5% or less in brain scan relevant geometries.
  • Combined scatter and attenuation correction demonstrated robust performance.

Conclusions:

  • The developed scatter correction algorithms are efficient and suitable for 3D PET brain imaging.
  • Accurate modeling and correction of scatter radiation improve quantitative accuracy in PET.
  • The proposed methods offer a practical solution for reducing scatter-induced artifacts in clinical PET studies.