Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Comparison between an automated and manual sphygmomanometer in a population survey
Martin G Myers1, Natalie H McInnis, George J Fodor
1Division of Cardiology, Schulich Heart Centre, Sunnybrook Health Sciences Centre, Department of Medicine, University of Toronto, Toronto, Ontario, Canada. martin.meyers@sunnybrook.ca
Background:
An automated sphygmomanometer, the BpTRU, was used in a blood pressure (BP) survey of 2,551 residents in the province of Ontario. Automated BP readings were compared with measurements taken by a mercury sphygmomanometer under standardized conditions in a random 10% sample.
Methods:
BP was recorded in 238 individuals in random order using both a standard mercury device and an automated BP recorder, the BpTRU. All subjects rested for 5 min prior to the first BP reading, which was then discarded. The mean of the next three readings was obtained using the mercury device whereas the BpTRU was set to record a mean of five readings taken at 1 min intervals with subjects resting alone in a quiet room.
Results:
The mean s.d. BP with the automated device was 115 +/- 16/71 +/- 10 mm Hg compared to 118 +/- 16/74 +/- 10 mm Hg for the manual BP (P < 0.001). A systolic BP > or = 140 mm Hg was present for 16 automated and 19 manual readings. Similarly, the diastolic BP was > or = 90 mm Hg for 9 automated and 14 manual readings. Linear regression analysis showed that automated BP was a significant (P < 0.001) predictor of both manual systolic and diastolic BP.
Conclusion:
Conventional manual BP readings can be replaced by readings taken using a validated, automated BP recorder in population surveys. The slightly lower readings obtained with the BpTRU device (in the context of reduced observer-subject interaction) may be a more accurate estimate of BP status.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure