Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Effect of echocardiographic contrast on velocity vector imaging myocardial tracking
Kwan S Lee1, Tadaaki Honda, Christina S Reuss
1Division of Cardiovascular Diseases, Mayo Clinic Arizona, Scottsdale, Arizona 85259, USA.
Background:
Two-dimensional speckle tracking is an evolving ultrasound technology that allows objective evaluation of left ventricular function. The effect of echocardiographic contrast with 2-dimensional speckle tracking image processing is poorly defined.
Methods:
A total of 11 patients undergoing clinically indicated transthoracic echocardiography who also required echocardiographic contrast for left ventricular endocardial border enhancement were prospectively studied. Offline velocity vector imaging analysis was performed to calculate global and regional longitudinal strain from the apical 4-chamber view while corresponding circumferential strain was obtained from the parasternal short-axis view at the papillary muscle level. Images were isochronally normalized and the systolic phase was studied precontrast and postcontrast opacification.
Results:
Intraclass correlation coefficients were high for global circumferential strain in the short-axis view (0.77) and global longitudinal strain in the apical 4-chamber view (0.87) precontrast and postcontrast. Significant regional variability was present, more pronounced in the apical 4-chamber view (intraclass correlation coefficient 0.24-0.64) versus short-axis view (intraclass correlation coefficient 0.64-0.68).
Conclusions:
Velocity vector imaging is able to quantify global and regional strain with and without contrast. There is, however, wide interindividual variability in regional strain quantification with and without contrast, potentially limiting the clinical applicability.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
12:12Echocardiographic Approaches and Protocols for Comprehensive Phenotypic Characterization of Valvular Heart Disease in Mice
Published on: February 14, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...