Ultrasound-based Pulse Wave Velocity Evaluation in Mice
Published on: February 14, 2017
Estimation of pulses in ultrasound B-scan images
1Electron. Inst., Tech. Univ. of Denmark, Lyngby.
Abstract:
It is shown, based on an expression for the received pressure field in pulsed medical ultrasound systems, that a common one-dimensional pulse can be estimated from individual A-lines. An autoregressive moving average (ARMA) model is suggested for the pulse, and an estimator based on the prediction error method is derived. The estimator is used on a segment of an A-line, assuming that the pulse does not change significantly inside the segment. Several examples of the use of the estimator on synthetic data measured from a tissue phantom and in vitro data measured from a calf's liver are given. They show that a pulse can be estimated even at moderate signal-to-noise ratios.