Related Experiment Video
Updated: Jul 7, 2026

Development of a 3D Graphene Electrode Dielectrophoretic Device
Published on: June 22, 2014
Design and construction of a preparative-scale dynamic field gradient focusing apparatus
Noah I Tracy1, Zheng Huang, Cornelius F Ivory
1School of Chemical Engineering and Bioengineering, Washington State University, Pullman, WA 99164, USA.
Abstract:
A linear model is used to show that dynamic field gradient focusing (DFGF) can be scaled to preparative capacity, approximately O (10 mgs). This paper explains how the preparative-scale DFGF apparatus was designed and fabricated. Scaled-down experiments and mathematical modeling guided material selection and design changes during construction to increase the probability that the prototype preparative-scale DFGF apparatus would perform as intended. The finished prototype successfully focused bovine hemoglobin from an initial concentration of 6.82 to 15 mg/mL and allowed for 86% recovery of injected protein.