Are melodic medical equipment alarms easily learned?
Alexandra N Wee1, Penelope M Sanderson
1School of Information Technology Electrical Engineering, University of Queensland, St. Lucia, Queensland, Australia.
Background:
We tested melodic auditory alarms recommended in the IEC 60601-1-8 standard for medical electrical equipment for ease of learning and discrimination, and for effectiveness during a timeshared task.
Methods:
Twenty-two critical care nurses learned the IEC 60601-1-8 melodic alarms over two training sessions more than a week apart, with or without mnemonics suggested in the standard. Subsequently, the nurses identified alarms arriving at quasi random intervals while performing a timeshared arithmetic task.
Results:
Only one nurse (4.5%) identified the alarms with 100% accuracy after two training sessions. Mnemonics did not improve overall alarm identification accuracy (mnemonic = 56%, nonmnemonic = 55%) but led to a narrower range of confusions between alarms. Nurses responded faster (P < 0.0001) and more accurately (P = 0.032) to medium priority than high priority alarms, despite rating high priority alarms as sounding more urgent (P < 0.0001). Nurses with at least 1 yr of formal musical training identified the alarms much more accurately (musical training = 73%, no musical training = 38%, P < 0.0001), perceived a greater distinction between high and medium priority alarms (P = 0.002), and found identifying the alarms easier overall (P = 0.023). During the timeshared task, nurses' responses were slower (P = 0.002) and became less accurate (P = 0.02).
Conclusions:
The slow rate of learning and persistent confusions suggest that the IEC 60601-1-8 melodic alarms should be redesigned before they are adopted for clinical practice.
More Related Videos
10:50Behavioral Determination of Stimulus Pair Discrimination of Auditory Acoustic and Electrical Stimuli Using a Classical Conditioning and Heart-rate Approach
Published on: June 6, 2012
11:50Clinical Practice Protocol of Creative Music Therapy for Preterm Infants and Their Parents in the Neonatal Intensive Care Unit
Published on: January 7, 2020
Related Concept Videos
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Operational Amplifiers