The interobserver and test-retest variability of the dysphonia severity index
Marieke M Hakkesteegt1, Marjan H Wieringa, Michael P Brocaar
1Department of Otorhinolaryngology, Erasmus MC-University Medical Center Rotterdam, The Netherlands.
Objective:
The purpose of this study was to investigate the interobserver variability and the test-retest variability of the Dysphonia Severity Index (DSI), a multiparametric instrument to assess voice quality.
Methods:
The DSI was measured in 30 nonsmoking volunteers without voice complaints or voice disorders by two speech pathologists. The subjects were measured on 3 different days, with an interval of 1 week.
Results:
The difference in DSI between two observers (interobserver difference) was not significant. The intraclass correlation coefficient for the DSI was 0.79. The standard deviation of the difference between two duplicate measurements by different observers was 1.27.
Conclusion:
Differences in measurements between different observers were not significant. The intraclass correlation coefficient of the DSI was 0.79, which is to be considered good. Differences in DSI within one patient need to be larger than 2.49 to be significant.
Related Concept Videos
Comparing Experimental Results: Student's t-Test
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.