Comparing the old and new generation SELDI-TOF MS: implications for serum protein profiling
Marie-Christine W Gast1, Judith Ymn Engwegen, Jan Hm Schellens
1Department of Pharmacy & Pharmacology, The Netherlands Cancer Institute/Slotervaart Hospital, Amsterdam, The Netherlands. Marie-Christine.Gast@slz.nl
The new PCS 4000 series Surface-Enhanced Laser Desorption/Ionization-Time of Flight Mass Spectrometry (SELDI-TOF MS) instrument showed no improvement over the PBS-IIc for crude serum analysis. However, it demonstrated superior performance in detecting low-abundant proteins after serum fractionation.
Area of Science:
- Biomarker Discovery
- Mass Spectrometry
- Proteomics
Background:
- The PBS-IIc SELDI-TOF MS is widely used for biomarker discovery but faces limitations in quantification and reproducibility.
- A new generation instrument, the PCS 4000 series, was developed with enhanced features to address these issues.
- This study evaluates the performance of the PBS-IIc and PCS 4000 series SELDI-TOF MS instruments.
Purpose of the Study:
- To compare the analytical performance of the PBS-IIc and PCS 4000 series SELDI-TOF MS instruments.
- To assess improvements in peak detection, biomarker potential, and reproducibility.
- To evaluate performance specifically after serum fractionation for low-abundant protein detection.
Main Methods:
- Two sample sets were analyzed using both PBS-IIc and PCS 4000 series instruments on the same ProteinChip arrays.
- Generated spectra were analyzed using associated software packages.
- Performance was evaluated based on peak detection, biomarker potential, reproducibility, and analysis of serum fractions.
Main Results:
- The PCS 4000 series instrument did not show improved performance in terms of peak detection, biomarker potential, or reproducibility for crude serum analysis compared to the PBS-IIc.
- However, the PCS 4000 series demonstrated superior performance in peak detection when analyzing serum fractions.
Conclusions:
- The PCS 4000 series SELDI-TOF MS instrument offers no advantage over the PBS-IIc for crude serum analysis.
- Its enhanced capability in analyzing serum fractions holds promise for improved detection and identification of low-abundant protein biomarkers.
- Serum fractionation is a valuable technique for reducing dynamic range and facilitating biomarker discovery.