ALC crystal oscillators based pressure and temperature measurement integrated circuit for high temperature oil well
N A Bianchi1, J M Karam, B Courtois
1TIMA Lab., Grenoble.
Abstract:
An application specific integrated circuit for the pressure and temperature measurement at the high temperature oil well bottom conditions is presented in this paper. The circuit is mainly composed of three high performance automatic level controlled (ALC) oscillator circuits, which drive three external crystals (the sensitive elements), as well as mixing and filtering stages. The integrated circuit was successfully tested up to 220 degrees C, showing a frequency resolution of 0.0003 ppm (0.0007 psi, for the pressure measurement) and a drift of 1.5 Hz/month (0.5 psi/month) concerning, respectively, the short- and long-term measurement stability. A drastic reduction of the whole measurement tool size and cost was then allowed by means of this application specific BiCMOS integrated circuit.
More Related Videos
09:46Adsorption Device Based on a Langatate Crystal Microbalance for High Temperature High Pressure Gas Adsorption in Zeolite H-ZSM-5
Published on: August 25, 2016
07:48An Externally-Heated Diamond Anvil Cell for Synthesis and Single-Crystal Elasticity Determination of Ice-VII at High Pressure-Temperature Conditions
Published on: June 18, 2020
Related Concept Videos
Constant Pressure Calorimetry
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pressure Gauges