Radiation exposure during elective coronary angioplasty: the effect of flat-panel detection
Ananth M Prasan1, Glen Ison, David M Rees
1Department of Cardiology St George Hospital, Level 1, Clinical Services Building, Gray Street, Kogarah, New South Wales 2217, Australia. Prasana@sesahs.nsw.gov.au
Background:
Coronary angiography and angioplasty have to date been performed using digital angiography and fluoroscopic systems which incorporate an image intensifier (II). More recently flat-panel (FP) detectors have been introduced which are thought to improve spatial resolution. However, there is limited data on the effect of flat-panel detection on radiation exposure. We sought to determine the impact of flat-panel on cumulative radiation exposure in patients undergoing elective coronary angioplasty at our institution.
Methods:
Patients who underwent elective coronary angioplasty in the six months prior to and following upgrade of our Toshiba catheterisation laboratory from image intensifier to flat-panel were included. Demographic and radiation data were collected prospectively and the same five operators performed interventions during the 12-month period. Radiation data was obtained from the dose-area product meter intrinsic to the fluoroscopy system.
Results:
One hundred and thirty seven patients underwent elective angioplasty over the 12-month period (68 II, 69 FP). Cumulative radiation exposure was increased in flat-panel cases (99, 129 Gy cm(2) versus 71, 77 Gy cm(2), p=0.001). This increase was independent of patient weight (78+/-15 kg versus 78+/-17 kg, p=NS), screening time (19+/-12 min versus 18+/-13 min, p=NS) and total number of digital acquisitions (1475, 820 versus 1668, 1365, p=NS). The total amount of contrast dye did not differ between flat-panel and image intensifier cases (195+/-76 ml versus 194+/-79 ml, p=NS).
Conclusions:
Adoption of flat-panel detector technology increases radiation exposure. This may have important safety implications for catheterisation laboratory staff and patients undergoing multiple interventional procedures.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation I: X-ray and CT
Cardiac Catheterization IV: Nursing Management