Related Experiment Video
Updated: Jul 7, 2026

Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
A fast response hafnium selective polymeric membrane electrode based on
B Rezaei1, S Meghdadi, R Fazel Zarandi
1Department of Chemistry, Isfahan University of Technology, Isfahan 84156-83111, Iran. rezaei@cc.iut.ac.ir
Abstract:
In this study a new hafnium selective sensor was fabricated from polyvinylchloride (PVC) matrix membrane containing neutral carrier N,N'-bis(alpha-methyl-salicylidene)-dipropylenetriamine (Mesaldpt) as a new ionophore, sodium tetraphenyl borate (NaTPB) as anionic discriminator and dioctyl phthalate (DOP) as plasticizing solvent mediator in tetrahydrofuran solvent. The electrode exhibits Nernstian response for Hf(4+) (Hafnium(IV)) over a wide concentration range (2.0 x 10(-7) to 1.0 x 10(-1)M) with the determination coefficient of 0.9966 and slope of 15.1+/-0.1 mVdecades(-1). The limit of detection is 1.9 x 10(-7)M. The electrode has a fast response time of 18s and a working pH range of 4-8. The proposed membrane shows excellent discriminating ability towards Hf(4+) ion with regard to several alkali, alkaline earth transition and heavy metal ions. It can be used over a period of 1.5 months with good reproducibility. It is successfully applied for direct determination of Hf(4+) in solutions by standard addition method for real sample analysis.