Open-tubular gas chromatography using capillary coated with octadecylamine-capped gold nanoparticles
Qi-Shu Qu1, Fei Shen, Ming Shen
1Jiangsu Key Laboratory of Environmental Materials and Environmental Engineering, College of Chemistry and Chemical Engineering, Yangzhou University, Yangzhou 225002, China. quqishu@gmail.com
Abstract:
The octadecylamine-capped gold nanoparticles (ODA-Au-NPs) were prepared and directly used to coat the capillary wall. The hydrophobic coating acted as the stationary phase for open-tubular gas chromatography (OTGC). The ODA-Au-NPs can be adsorbed tightly onto the inner surface of fused silica capillary column via electrostatic interaction and enhanced interaction of van der Waals between gold nanoparticles and the capillary wall. Thus, the modification of the inner surface of capillary column by ODA-Au-NPs can be achieved simply by flushing the capillary with a solution of ODA-Au-NPs and the resulted ODA-Au-NPs coating is very stable. No perceptible degradation in the ODA-Au-NPs-based separation was observed after approximately 1900 sample runs. This type of columns also provided excellent chromatographic performances: high number of theoretical plates, outstanding run-to-run and column-to-column reproducibility, and high selectivity for a wide range of test mixtures. An efficiency of 2474 theoretical plates per meter for chlorobenzene was obtained on an ODA-Au-NPs-modified 1.6 m x 100 microm i.d. fused silica capillary column.
More Related Videos
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
07:49On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...