A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
Imaging Performance of a Dynamic Positron Emission Tomograph: Positome IIIp
Abstract:
This paper documents modifications to an older PET system to improve its dead time, scatter fraction, and spatial resolution in high count rate, short duration studies. A new dual-tapered collimator reduces scatter by 33 percent while providing excellent resolution uniformity in all slices. A data encoding scheme produces uniformly sampled parallel projections from the coincidence data in real time while the detector array executes an orbital motion. The image uniformity, scatter compensation and high count-rate performance have been validated up to 40 kBq/cc in a 20 cm flood source. The errors in image quantification due to counting statistics, live time, and random counts are estimated from repeated measurements on a contrast phantom at high count-rates. The effects of two methods of scatter compensation on image contrast are shown in contrast phantoms and a typical glucose utilization study. Blood flow measurements using 0-15 labeled water bolus method, made under different physiological conditions, reflected the changes expected. The true count efficiency of 75 kcps/(uCi/cc) permits these studies to be done with only 500-900 MBq (13-25 mCi) injected activity.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...