An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Simulation and experimental characterization of a 2-D capacitive micromachined ultrasonic transducer array element
O Oralkan1, X Jin, F L Degertekin
1Edward L. Ginzton Lab., Stanford Univ., CA.
Abstract:
In this letter, a 400-mumx400-mum 2-D capacitive micromachined ultrasonic transducer (cMUT) array element is experimentally characterised, and the results are found to be in good agreement with theoretical predictions. As a receiver, the transducer has a 1.8x10(-7) nm/ radical(Hz) displacement sensitivity, and, as a transmitter, it produces 16.4 kPa/V of output pressure at the transducer surface at 3 MHz. The transducer also has more than 100% fractional bandwidth around 3 MHz, which makes it suitable for ultrasound imaging.