Katja Seme1, Tina Mocilnik, Kristina Fujs Komlos
1Institute of Microbiology and Immunology, Faculty of Medicine, Zaloska 4, Ljubljana 1000, Slovenia. mario.poljak@mf.uni-lj.si.
Abstract:
A prospective unblinded comparative evaluation of three assays for the detection of enteroviral RNA performed on 83 positive and 79 negative cerebrospinal fluid samples showed initial and resolved sensitivities of 90.4% and 98.8%, respectively, for the Cepheid GeneXpert enterovirus assay; 94.0% and 97.6%, respectively, for the Argene enterovirus consensus kit; and 100% and 100%, respectively, for an in-house real-time PCR. The initial and resolved specificities were 100% for all three assays.
More Related Videos
12:32EPA Method 1615. Measurement of Enterovirus and Norovirus Occurrence in Water by Culture and RT-qPCR. Part III. Virus Detection by RT-qPCR
Published on: January 16, 2016
10:48EPA Method 1615. Measurement of Enterovirus and Norovirus Occurrence in Water by Culture and RT-qPCR. I. Collection of Virus Samples