Breakage of a volar locking plate after delayed union of a distal radius fracture
Tom De Baere1, Frédéric Lecouvet, Olivier Barbier
1Cliniques Universitaires St-Luc, Bruxelles, Belgium.
Abstract:
This is a case report of a 58-year-old obese, diabetic female who sustained a dorsally angulated distal radius fracture, treated with a volar locking plate. Union was delayed and the plate broke. The fracture healed after repeated surgery involving replacement of the plate and grafting with demineralised bone matrix. Delayed unions of distal radius fractures are rare. This case shows that fracture treatment with locking compression plates has some specific biomechanical consequences which have to be taken into account, and special attention will be given to technical peculiarities and possible drawbacks.
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...