Can Modern Infrared Analyzers Replace Gas Chromatography to Measure Anesthetic Vapor Concentrations?
Jan Fa Hendrickx1, Hendrikus Jm Lemmens, Rik Carette
1Department of Anesthesia, Stanford University School of Medicine, Stanford, California, USA. jcnwahendrickx@yahoo.com.
Background:
Gas chromatography (GC) has often been considered the most accurate method to measure the concentration of inhaled anesthetic vapors. However, infrared (IR) gas analysis has become the clinically preferred monitoring technique because it provides continuous data, is less expensive and more practical, and is readily available. We examined the accuracy of a modern IR analyzer (M-CAiOV compact gas IR analyzer (General Electric, Helsinki, Finland) by comparing its performance with GC.
Methods:
To examine linearity, we analyzed 3 different concentrations of 3 different agents in O₂: 0.3, 0.7, and 1.2% isoflurane; 0.5, 1, and 2% sevoflurane; and 1, 3, and 6% desflurane. To examine the effect of carrier gas composition, we prepared mixtures of 1% isoflurane, 1 or 2% sevoflurane, or 6% desflurane in 100% O₂ (= O₂ group); 30%O₂+ 70%N₂O (= N₂O group), 28%O₂ + 66%N₂O + 5%CO₂ (= CO₂ group), or air. To examine consistency between analyzers, four different M-CAiOV analyzers were tested.
Results:
The IR analyzer response in O2 is linear over the concentration range studied: IR isoflurane % = -0.0256 + (1.006 * GC %), R = 0.998; IR sevoflurane % = -0.008 + (0.946 * GC %), R = 0.993; and IR desflurane % = 0.256 + (0.919 * GC %), R = 0.998. The deviation from GC calculated as (100*(IR-GC)/GC), in %) ranged from -11 to 11% for the medium and higher concentrations, and from -20 to +20% for the lowest concentrations. No carrier gas effect could be detected. Individual modules differed in their accuracy (p = 0.004), with differences between analyzers mounting up to 12% of the medium and highest concentrations and up to 25% of the lowest agent concentrations.
Conclusion:
M-CAiOV compact gas IR analyzers are well compensated for carrier gas cross-sensitivity and are linear over the range of concentrations studied. IR and GC cannot be used interchangeably, because the deviations between GC and IR mount up to ± 20%, and because individual analyzers differ unpredictably in their performance.
Related Concept Videos
Inhalational Anesthetics: Overview
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-II
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...