Related Experiment Video
Updated: Jul 7, 2026

Wideband Optical Detector of Ultrasound for Medical Imaging Applications
Published on: May 11, 2014
The design of protection circuitry for high-frequency ultrasound imaging systems
G R Lockwood1, J W Hunt, F S Foster
1Dept. of Med. Biophys., Toronto Univ., Ont.
Abstract:
Transmission line lengths in the protection circuitry of a high-frequency (>20-MHz) ultrasound imaging system have an important effect on the frequency, amplitude, and bandwidth of the pulse-echo response of the system. A model that includes the transmission line lengths between the pulser, transducer, and receiver and the electromechanical properties of high-frequency transducers is used to illustrate the importance of correctly choosing these line lengths. An iterative optimization procedure for designing the protection circuitry for a broadband system is proposed. A theoretical and experimental analysis of the validity of this approach is reported for a 45-MHz PVDF transducer.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Design Example
Radial System Protection
In a radial system with a fault downstream of the third breaker, ideally, only the third breaker will open, isolating the fault and interrupting the load connected beyond it. The second breaker has a longer delay setting,...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Line Protection with Impedance Relays