Fracturing and stretching of angioplasty balloon catheters under bending with tension
W D Jeans1, A Blakeborough, H W Loose
1Department of Radiodiagnosis, University of Bristol, UK.
Abstract:
Eight catheters from five manufacturers have been subjected to tests which combined bending and tension. Three failure mechanisms have been identified. These included separation of the balloon from its proximal attachment to the shaft, fracture of the shaft at its junction with the balloon, and undue ductility of the shaft. It is concluded that good catheter design should ensure that there is no sudden change in stiffness along the length of the catheter, and that the optimum relationship of ductility to stiffness has not yet been defined for catheters. Recommendations relating to these characteristics could be included in the British Standards relating to angioplasty balloon catheters.