Iatrogenic vasospasm in carotid artery stent angioplasty with distal protection devices
Clemens M Schirmer1, Daniel A Hoit, Adel M Malek
1Cerebrovascular and Endovascular Division, Department of Neurosurgery, Tufts-New England Medical Center and Tufts University School of Medicine, Boston, Massachusetts 02111, USA.
Insights
First-generation FilterWire EX distal protection devices (DPDs) significantly increased vasospasm during carotid artery stenting. Second-generation mobile-basket DPDs and no protection showed lower vasospasm rates, suggesting improved safety.
Area of Science:
- Interventional Cardiology
- Vascular Surgery
- Medical Device Technology
Background:
- Distal protection devices (DPDs) reduce embolic stroke risk in carotid artery (CA) stenting.
- First-generation FilterWire EX has a fixed-basket design.
- Second-generation mobile-basket DPDs offer greater flexibility.
Purpose of the Study:
- Compare vasospasm incidence with fixed-basket DPDs, mobile-basket DPDs, and no protection during CA stenting.
- Evaluate the safety and efficacy of different DPDs in preventing procedural complications.
Main Methods:
- Retrospective analysis of 40 CA stent placements (37 patients).
- Procedures utilized fixed-basket FilterWire EX DPD (n=12), second-generation mobile-basket DPD (n=6), or no protection (n=22).
- Recorded and analyzed clinical presentation, angiographic findings (vasospasm, dissection), and outcomes.
Main Results:
- High vasospasm rates with fixed-basket FilterWire EX (67%) compared to mobile-basket DPDs (17%) and no protection (14%).
- FilterWire EX use was an independent risk factor for vasospasm (p < 0.0003).
- Secondary angioplasty also correlated with intraprocedural vasospasm.
Conclusions:
- Fixed-basket FilterWire EX DPD is associated with a significantly higher incidence of vasospasm during CA stenting.
- Second-generation mobile-basket DPDs did not increase vasospasm risk compared to unprotected procedures.
- Recommend using second-generation mobile-basket DPDs to minimize embolic and iatrogenic vascular injury risks.
Object:
Distal protection devices (DPDs) have decreased the risk of embolic stroke among patients with carotid artery (CA) disease undergoing CA stent placement. The FilterWire EX is a first-generation fixed-basket DPD with a filter rigidly attached to a guidewire. Second-generation mobile-basket DPDs (RX Accunet or SpiderFX) allow movement of the filter relative to the guidewire and can thus reduce the potential for vessel irritation, vasospasm, or intimal injury during CA stent placement.
Methods:
Stent angioplasty was attempted in 40 CAs (37 patients) using the fixed-basket FilterWire DPD, a second-generation mobile-basket DPD, or no protection in 12, 6, and 22 arteries, respectively. Clinical presentation, angiographic details relating to the incidence of vasospasm or dissection, and clinical outcome data were recorded and analyzed.
Results:
Vasospasm was associated with use of the fixed-basket FilterWire device (8 [67%] of 12 cases) compared with the bare unfiltered guidewire group (3 [14%] of 22) and the second-generation mobile-basket DPD group (1 [17%] of 6, p < 0.004). Secondary angioplasty was also associated with intraprocedural vasospasm. In a multivariate analysis, FilterWire use was an independent risk factor for vasospasm (p < 0.0003).
Conclusions:
A high incidence of vasospasm was observed following CA stent placement procedures in which the fixed-basket FilterWire EX DPD was used but not in unprotected CA stent placement or procedures in which a second-generation mobile-basket DPD was used. Although this phenomenon was self-limited in all instances, vasospasm should be considered a risk of these devices and may predispose to more serious vascular injury. Coronary artery stent placement should be performed with a second-generation mobile-basket DPD to minimize the risks of embolic complications and iatrogenic vascular injury.