Dislocation of rotating-hinge knee prostheses with antidislocation mechanism.

Daniel Pacha-Vicente1, Amer Malik, Enric Castellet-Feliu

  • 1Department of Orthopaedic Surgery, Vall d'Hebron Hospital, Barcelona, Spain.

The Journal of Arthroplasty
|February 19, 2008
PubMed
Summary

Two cases of femorotibial dislocation occurred in 192 Endo-Model knee reconstructions. This complication arose from antidislocation feature loosening, leading to excessive laxity and implant dislocation in rotating-hinge prostheses.

Related Concept Videos

Knee Joint01:23

Knee Joint

The knee joint is the most complicated joint in the body. It consists of three articulations– two tibiofemoral and one patellofemoral. As is characteristic of synovial joints, the knee joint has a thin articular capsule that partially surrounds this joint cavity. Additionally, several ligaments, muscles, and cartilaginous structures support the movement of the knee.
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris group...
Bones of the Lower Limb: Femur and Patella01:16

Bones of the Lower Limb: Femur and Patella

The femur is the body's longest and strongest bone spanning the thigh region. Its head articulates with the acetabulum of the hip bone to form the hip joint. A minor indentation on the medial side of the femoral head, called the fovea capitis, serves as the site of attachment for the ligament of the head of the femur. This weak ligament spans the femur and acetabulum and supports the hip joint. The narrowed region below the head is the neck of the femur. The inclination angle between the neck...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Society of Vascular and Interventional Neurology Standards and Parameters for Guideline Development and Publication.

Stroke (Hoboken, N.J.)·2026
Same author

Garceau Technique Requires Less Mechanical Work Than Modified Ponseti for Tibialis Anterior Tendon Transfer: A Biomechanical Study in Cadavers.

Clinical orthopaedics and related research·2025
Same author

Endovascular thrombectomy for large vessel occlusion in acute ischemic stroke patients with concomitant intracranial hemorrhage.

Journal of clinical neuroscience : official journal of the Neurosurgical Society of Australasia·2025
Same author

Endovascular Treatment of Patients With Acute Ischemic Stroke With Tandem Lesions Presenting With Low Alberta Stroke Program Early Computed Tomography Score.

Journal of the American Heart Association·2024
Same author

Anesthetic management for large vessel occlusion acute ischemic stroke with tandem lesions.

Journal of neurointerventional surgery·2024
Same author

Acromial physeal fracture in an adolescent: a case report and literature review.

JSES international·2024

Area of Science:

  • Orthopedic surgery
  • Biomedical engineering
  • Prosthetics and implants

Background:

  • The Endo-Model rotating-hinge prosthesis features an antidislocation mechanism designed to prevent dislocation.
  • Difficult knee reconstructions often utilize such advanced prosthetic designs.

Observation:

  • A review of 192 consecutive Endo-Model prosthesis implantations identified two instances of femorotibial dislocation.
  • These dislocations occurred despite the presence of the antidislocation feature.

Findings:

  • Loosening of the antidislocation feature can lead to increased flexion laxity.
  • This laxity permits excessive distraction, ultimately causing implant dislocation.

Implications:

  • This study highlights a potential failure mode in rotating-hinge knee prostheses with antidislocation designs.

Related Experiment Videos

  • Further research may be needed to refine or reinforce antidislocation mechanisms in knee implants.
  • Clinicians should be aware of this potential complication during difficult knee reconstructions.