Prediction of perceptible artifacts in JPEG2000 compressed abdomen CT images using a perceptual image quality metric
Bohyoung Kim1, Kyoung Ho Lee, Kil Joong Kim
1Department of Radiology, Seoul National University Bundang Hospital, 300 Gumi-dong, Bundang-gu, Seongnam-si, Gyeonggi-do, 463-707, Korea.
Rationale And Objectives:
To test a perceptual quality metric (high-dynamic range visual difference predictor, HDR-VDP) in predicting perceptible artifacts in Joint Photographic Experts Group 2000 compressed thin- and thick-section abdomen computed tomography images.
Materials And Methods:
A total of 120 thin (0.67 mm) and corresponding thick (5 mm) sections were compressed to ratios from 4:1 to 15:1. Peak signal-to-noise ratio (PSNR), HDR-VDP results (paired t-tests), and five radiologists' pooled responses for the presence of artifacts (exact tests for paired proportions) were compared between the thin and thick sections. For three subsets of 120 thin- (subset A), 120 thick- (subset B), and 60 thin- and 60 thick-section compressed images (subset C), receiver operating curve analysis was performed to compare PSNR and HDR-VDP in predicting the radiologists' responses. Using the cutoff values where the sum of sensitivity and specificity was the maximum in subset C, visually lossless thresholds (VLTs) were estimated for the 240 original images and the estimation accuracy was compared (McNemar test).
Results:
Thin sections showed more artifacts in terms of PSNR, HDR-VDP, and radiologists' responses (p < .0001). HDR-VDP outperformed PSNR for subset C (area under the curve: 0.97 versus 0.93, p = 0.03), whereas they did not differ significantly for subset A or B. Using the cutoff values, PSNR and HDR-VDP predicted the VLT accurately for 124 (51.7%) and 183 (76.3%) images, respectively (p < .0001).
Conclusions:
HDR-VDP can predict the perceptible compression artifacts, and therefore can be potentially used to estimate the VLT for such compressions.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Ultrasonography
During an ultrasonography procedure, a handheld device called a...