Radiation dose survey in a paediatric cardiac catheterisation laboratory equipped with flat-panel detectors
O Dragusin1, M Gewillig, W Desmet
1Department of Radiology, University Hospital Gasthuisberg, 49 Herestraat, B-3000 Leuven, Belgium. odragusin@yahoo.com
Insights
This study assessed radiation doses from new flat-panel X-ray systems in pediatric cardiac catheterizations. Results provide crucial data for optimizing radiation safety in young patients undergoing these procedures.
Area of Science:
- Medical Imaging
- Pediatric Cardiology
- Radiation Physics
Background:
- Flat-panel X-ray detectors are increasingly used in pediatric cardiac catheterization.
- Infants and children are particularly sensitive to radiation exposure.
- Understanding radiation doses is critical for risk management in pediatric procedures.
Purpose of the Study:
- To investigate and report radiation doses delivered by new flat-panel X-ray systems during pediatric cardiac catheterizations.
- To establish dose-area product (DAP) and effective dose (E) parameters for different age groups.
- To provide data for optimizing radiation protection in pediatric fluoroscopy.
Main Methods:
- A survey of 273 pediatric cardiac catheterizations (diagnostic and therapeutic) was conducted.
- Radiation doses were analyzed using dose-area product (DAP) and fluoroscopy time.
- Effective dose (E) was calculated using PCXMC software for risk estimation across six age groups.
- Dose conversion factors relating DAP and E were estimated for each age group.
Main Results:
- Effective dose (E) varied by age and procedure type, with higher doses for therapeutic interventions.
- For diagnostic procedures, the third quartile of E ranged from 11.3 mSv (newborns) to 7 mSv (10-15 years).
- For therapeutic procedures, the third quartile of E ranged from 22.6 mSv (0-30 days) to 34.1 mSv (10-15 years).
Conclusions:
- This study establishes baseline radiation dose metrics for pediatric cardiac catheterizations using modern flat-panel X-ray systems.
- The findings are essential for initiating radiation dose optimization strategies.
- Further utilization of dose reduction potentials of flat-panel systems is recommended for pediatric patient safety.
Abstract:
Flat-panel X-ray detectors for fluoroscopy represent a modern imaging equipment that is being implemented in paediatric cardiac catheterisation laboratories. Infants and children represent a group of patients with a high radiosensitivity. A survey of 273 (126 diagnostic and 147 therapeutic) paediatric catheterisations was performed to investigate the radiation doses delivered by the new X-ray system. Statistical parameters (75th, 50th and 25th percentiles) of dose-area product (DAP) and fluoroscopy time are reported for patients divided into six age groups: 0-30 d, >1-12 m, >1-3, >3-5, >5-10 and >10-15 y. For accurate risk estimation, effective dose (E) has been determined for all patients using the PCXMC software. For diagnostic procedures, the third quartile of E ranges from 11.3 mSv for newborns to 7 mSv for children of 10-15 y. Therapeutic procedures are more complex than diagnostic. Consequently, the third quartile of E is 22.6 mSv (0-30 d), 18.6 (>1-12 m), 13.3 (>1-3 y), 21.5 (>3-5 y), 17.8 (>5-10 y) and 34.1 mSv (>10-15 y). Dose conversion factors, which relate the DAP and E, have been estimated for each age group. The results of this study may serve as a first step in the optimisation process, in order to make full use of the dose reduction potential of flat-panel systems.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation