Threshold dosages for damage to mammalian liver by high intensity focused ultrasound
1Dept. of Electr. and Comput. Eng., Illinois Univ., Urbana, IL.
Abstract:
The threshold dosages for high-intensity focal lesion production were determined at 3 MHz in the cat liver for exposure durations covering the range 0.003-35 s. The liver threshold was found to parallel that for the brain over the range of exposure durations 0.1 to 10 s, but to be more than twice the intensity level, approximately following the relation IT(0.5)=460 W-cm(2)/s(2), where I is the peak intensity and T is the exposure duration. At shorter exposure durations the threshold curve deviates from this relation, probably due to a change to a transient cavitation mechanism of damage.
More Related Videos
14:10Contrast Ultrasound Targeted Treatment of Gliomas in Mice via Drug-Bearing Nanoparticle Delivery and Microvascular Ablation
Published on: December 15, 2010
09:05Generation and Quantitative Analysis of Pulsed Low Frequency Ultrasound to Determine the Sonic Sensitivity of Untreated and Treated Neoplastic Cells
Published on: July 22, 2015