Sixteen-slice multidetector computed tomography for graft patency evaluation after coronary artery bypass surgery
Miguel Sousa Uva1, Fernando Matias, Armindo Mesquita
1Division of Cardiac Surgery, Hospital da Cruz Vermelha Portuguesa, Lisbon, Portugal. miguel.uva@hcvp.com.pt
Objective:
To investigate the ability of 16-slice multidetector computed tomography (MDCT) to assess coronary artery bypass graft patency and to detect bypass stenosis by comparison with coronary angiography.
Methods:
Thirty patients underwent both conventional coronary bypass angiography and retrospective ECG gated 16-slice multidetector computed tomography after surgery using 0.4 seconds rotation time and 1.25 mm slice thickness.
Results:
Among a total of 107 bypass grafts, 101 grafts (94.4%) were evaluable by MDCT. Thirteen patients were taking oral beta blockers. Mean heart rate was 73.6 (52-105). Of the 40 internal mammary arteries and two radial arteries examined, only one was occluded by coronary bypass angiography and MDCT, resulting in a sensitivity of 100% and a specificity of 100%. MDCT correctly diagnosed all patent venous grafts and missed two of the 14 venous grafts shown occluded by conventional angiography resulting in a sensitivity of 85.7% and specificity of 100%. When occluded grafts were excluded, MDCT did not detect two out of two anastomotic arterial graft stenosis >50% and resulted in one false positive result for a sensitivity and specificity of 0% and 97.4%, respectively. MDCT correctly diagnosed one out of three venous stenosis >50% and falsely diagnosed one venous graft stenosis >50% yielding a 33.3% and 97.6% sensitivity and specificity, respectively.
Conclusion:
Sixteen-slice MDCT allows for noninvasive evaluation of coronary bypass grafts patency with high diagnostic accuracy. Assessment of distal anastomotic stenosis was deficient, particularly for arterial grafts, still limited by low resolution or artifacts. Improved accuracy may be obtained by more aggressive heart rate reduction.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
05:25Surgical Porcine Model of Chronic Myocardial Ischemia Treated by Exosome-laden Collagen Patch and Off-pump Coronary Artery Bypass Graft
Published on: September 15, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Acute Coronary Syndrome III: Diagnostic Studies
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...