Cardiac output derived from arterial pressure waveform analysis in patients undergoing cardiac surgery: validity of a
Jochen Mayer1, Joachim Boldt, Michael W Wolf
1Department of Anesthesiology and Intensive Care Medicine, Klinikum Ludwigshafen, Bremserstr. 79, 67063 Ludwigshafen, Germany. j-mayer@gmx.de
Background:
The performance of a recently introduced, arterial waveform-based device for measuring cardiac output (CO) without the need of invasive calibration (FloTrac/Vigileo) has been controversial. We designed the present study to assess the validity of an improved version of this monitoring technique compared with intermittent thermodilution CO measurement using a pulmonary artery catheter in patients undergoing cardiac surgery.
Methods:
Forty ASA III patients scheduled for elective coronary artery bypass grafting with cardiopulmonary bypass (CPB) were studied. Simultaneous CO measurements by bolus thermodilution and the FloTrac/Vigileo device were obtained after induction of anesthesia (T1), before CPB (T2), after CPB (T3), after sternal closure (T4), on arrival in the intensive care unit (T5), 4 h (T6), 8 h (T7), and 24 h after surgery (T8). CO was indexed to the body surface area (cardiac index, CI). A percentage error of 30% or less was established as the criterion for method interchangeability.
Results:
Two hundred and eighty-two data pairs were analyzed. Thermodilution CI ranged from 1.2 to 4.1 L x min(-1) x m(-2) (mean 2.5 +/- 0.54 L x min(-1) x m(-2)). Bias and precision (1.96 sd of the bias) were 0.19 L x min(-1) x m(-2) and +/- 0.60 L x min(-1) x m(-2), resulting in an overall percentage error of 24.6%. Subgroup analysis revealed a percentage error of 28.3% for data pairs obtained intraoperatively (T1-4) and 20.7% in intensive care unit (T5-8).
Conclusion:
CI values obtained by the improved, second generation semiinvasive arterial waveform device showed good intraoperative and postoperative agreement with intermittent pulmonary artery thermodilution CI measurements in patients undergoing coronary artery bypass graft surgery.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...